Accurate and reliable blood pressure measurement is critical for the proper diagnosis and management of hypertension. So much so that a 5 mm Hg BP measurement error can lead to incorrect hypertension classification in 84 million individuals worldwide, according to a consensus document published in the Journal of Hypertension. Understanding the ways BP measurement goes wrong, and how to tackle them, can improve diagnosis and management of hypertension.
Unfortunately, BP measurement is often suboptimally performed in clinical practice, which can lead to errors that inappropriately alter management decisions in 20% to 45% of cases. This inaccuracy has persisted despite extensive education and efforts to raise awareness on the adverse consequences of incorrect clinic BP measurement, according to the Lancet Commission on Hypertension Group position statement. That statement was co-written by AMA Group Vice President of Improving Health Outcomes Michael Rakotz, MD, and Gregory D. Wozniak, PhD, who is vice president of health outcomes analytics at the AMA.
“Many measurement errors can be minimized by appropriate patient preparation and standardized techniques. Validated semi-automated or automated upper arm cuff devices should be used instead of auscultation to simplify measurement and prevent observer error,” the consensus document says.
The consensus document cites multiple causes of inaccuracy in measurement. Here are four ways BP measurement goes wrong and how to address them.
Patient-related
There are instances in which the BP-measurement accuracy is affected by patients’ habits or behaviors (PDF). Acute meal ingestion, caffeine or nicotine use can all affect BP readings, leading to errors in measurement accuracy. If the patient has a full bladder, that can lead to an error in systolic BP of up to 33 mm Hg, and the white-coat effect can have an error of up to 26 mm Hg.
It is important for the patient to rest comfortably in a quiet environment for five minutes in a chair. The patient should also have an empty bladder and not have eaten, ingested caffeine, smoked or engaged in physical activity at least 30 minutes before the measurement.
Procedure-related
Inaccuracies in BP measurement can also occur due to errors related to the procedure. For example, having the patient’s arm lower than heart level can lead to an error of 4 mm Hg up to 23 mm Hg.
Procedure-related errors might also occur if the patient’s legs are crossed at the knees or if talking occurs during BP measurement. A fast cuff deflation rate also can contribute to inaccuracy.
AMA MAP™ Hypertension is an evidence-based quality improvement program that provides a clear path to significant, sustained improvements in BP control. With the AMA MAP program, health care organizations can increase BP-control rates quickly. The program has demonstrated a 10% increase in BP control in six months with sustained results at one year.
Equipment-related
If a cuff is too small or too large, errors in measurement can occur. Adding to inaccuracy are automated devices that have not been tested for accuracy, which can account for errors in systolic BP.
“An important issue with automated devices is that many have not been clinically validated for measurement accuracy,” says the statement. “Clinical validation involves demonstrating that the device meets the accuracy requirements of international BP measurement standards.”
The process of clinical validation involves performing a protocol-based comparison using multiple measurements against blinded, two-observer auscultatory reference standard. For greater accuracy, only validated devices should be used.
Recently revamped, the US Blood Pressure Validated Device Listing (VDL™) can be used to find blood pressure-measurement devices that have been validated for clinical accuracy and are available in the U.S. In 2020, 16 BP measurement devices met the requirements to be listed as clinically accurate after the testing data was submitted by their manufacturers for review.
Now that number has grown to 102 and includes home and in-office upper arm devices, wrist cuff devices, community kiosks and 24-hour ambulatory monitors. Discover why a validated device listing is important for BP control.
Physician- or health professional-related
One common error in the clinical setting is failure to include a five-minute rest period. Errors can also include talking during the measurement procedure, using an incorrect cuff size and failure to take multiple measurements.
Time constraints are also quite common for casual measurements. This is because a casual reading takes about two minutes to perform compared with eight minutes for a standardized measurement. Physician readings were also found to be higher than nurse readings, which is the white coat effect in action.
The physician, nurse or other health professional is responsible for performing proper BP measurement while ensuring—to the greatest extent possible—that all potential causes of inaccuracy are avoided.
Training programs can lead to short-term success in BP readings. These can be web-based or in-person. However, shorter web-based programs may be preferred because of their practical advantages, lower cost and scalability.
Quality improvement programs that combine use of automated office BP measurement with physician and care team education on proper measurement, as well as advice on clinical workflow enhancement, can also improve readings.
The CME series, “AMA Measure BP Accurately,” is enduring material and designated by the AMA for a total of 1.50 PRA Category 1 Credit™.
The series is part of the AMA Ed Hub™, an online learning platform that brings together high-quality CME, maintenance of certification, and educational content from trusted sources, all in one place—with activities relevant to you, automated credit tracking, and reporting for some states and specialty boards.
Learn more about AMA CME accreditation.